6: SCPI command reference Model 2461 Interactive SourceMeterĀ® Instrument
Reference Manual
6-184 2461-901-01 A/November 2015
:TRIGger:BLENder<n>:OVERrun?
This command indicates whether or not an event was ignored because of the event detector state.
Type Affected by Where saved Default value
Query only Not applicable Not applicable Not applicable
Usage
:TRIGger:BLENder<n>:OVERrun?
<n>
The blender number (1 or 2)
Details
Indicates if an event was ignored because the event detector was already in the detected state when
the event occurred. This is an indication of the state of the event detector that is built into the event
blender itself.
This command does not indicate if an overrun occurred in any other part of the trigger model or in any
other trigger object that is monitoring the event. It also is not an indication of an action overrun.
Example
:TRIG:BLEN1:OVER?
If an event was ignored, the output is 1.
If an event was not ignored, the output is
0
.
Also see
:TRIGger:BLENder<n>:CLEar (on page 6-182)
:TRIGger:BLENder<n>:STIMulus<m>
This command specifies the events that trigger the blender.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Trigger blender clear
Save settings
NONE
Usage
:TRIGger:BLENder<n>:STIMulus<m> <event>
:TRIGger:BLENder<n>:STIMulus<m>?
<n>
The blender number (1 or 2)
<m>
The stimulus input number (1 to 4)
<event>
See Details