Interactive SourceMeter® Instrument Reference Manual Section 6:
2461-901-01 A/November 2015 6-57
Example 1
CURR:AVER:TCON MOV
Set the averaging filter type to moving average,
with a filter count of 10.
Enable the averaging filter.
Example 2
RES:AVER:TCON MOV
Set the averaging filter type to moving average,
with a filter count of 10.
Enable the averaging filter.
Example 3
VOLT:AVER:TCON MOV
VOLT:AVER ON
For voltage measurements, set the averaging
filter type to moving average, with a filter count
of 10.
Enable the averaging filter.
Also see
Filtering measurement data (on page 4-23)
[:SENSe[1]]:<function>:AVERage:COUNt (on page 6-53)
[:SENSe[1]]:<function>:AVERage[:STATe] (on page 6-55)
[:SENSe[1]]:<function>:AZERo[:STATe]
This command enables or disables automatic updates to the internal reference measurements (autozero) of the
instrument.
Type Affected by Where saved Default value
Instrument reset
Power cycle
Measure configuration list
Measure configuration list
Usage
[:SENSe[1]]:<function>:AZERo[:STATe] <state>
[:SENSe[1]]:<function>:AZERo[:STATe]?
The measure function:
• Current: CURRent[:DC]
• Resistance: RESistance
• Voltage:
The status of autozero:
• Disable autozero: OFF or 0
•