EasyManua.ls Logo

Keithley 2461

Keithley 2461
1109 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
D: Convert Model 2400 to Model 2461 code Model 2461 Interactive SourceMeterĀ® Instrument
Reference Manual
D-28 2461-901-01 A/November 2015
Model 2400 command
:SYSTem:RWLock
:SYSTem:RWLock?
Model 2461 command
Not available
Notes
No RS-232 communications available in the Model 2461.
Model 2400 command
:SYSTem:TIME?
Model 2461 command
:SYSTem:TIME (on page 6-155)
:SYSTem:TIME?
Notes
Model 2461 uses absolute time.
Model 2400 command
:SYSTem:TIME:RESet
Model 2461 command
Not available
Notes
Model 2461 uses absolute time.
Model 2400 command
:SYSTem:TIME:RESet:AUTO
Model 2461 command
Not available
Notes
Model 2461 uses absolute time.
Model 2400 command
:SYSTem:VERSion?
Model 2461 command
:SYSTem:VERSion? (on page 6-156)
Notes
TRACe subsystem
Model 2400 command
:TRACe:CLEar
Model 2461 command
:TRACe:CLEar (on page 6-159)
Notes
Allows selection of the buffer to clear.
Model 2400 command
:TRACe:DATA?
Model 2461 command
:TRACe:DATA? (on page 6-160)
Notes
In the Model 2461, this command allows you to dynamically specify the buffer
elements to retrieve from the reading buffer.
Model 2400 command
:TRACe:FEED
:TRACe:FEED?
Model 2461 command
Not available
Notes
Model 2400 command
:TRACe:FEED:CONTrol
:TRACe:FEED:CONTrol?
Model 2461 command
Not available
Notes
Model 2400 command
:TRACe:FREE?
Model 2461 command
Not available
Notes

Table of Contents

Other manuals for Keithley 2461