Interactive SourceMeter® Instrument Reference Manual Section 8:
2461-901-01 A/November 2015 8-7
TSP commands
acal.count
This attribute returns the number of times automatic calibration has been run.
Type TSP-Link accessible Affected by Where saved Default value
Usage
value = acal.count
The number of times autocalibration has been run
Details
The number of times that autocalibration has been run since the last factory calibration. The count
restarts at 1 after a factory calibration.
Example
Returns the number of times autocalibration has been
run.
Example output:
Also see
Autocalibration (on page 3-51)
acal.run() (on page 8-9)
acal.lastrun.internaltemp
This attribute returns the internal temperature of the instrument when autocalibration was run.
Type TSP-Link accessible Affected by Where saved Default value
Usage
temperature = acal.lastrun.internaltemp
Details
The temperature is displayed in Celsius (°C).