Model 2461
Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2461-901-01 A/November 2015 6-163
Also see
:FORMat[:DATA] (on page 6-46)
Reading buffers (on page 3-2)
Remote buffer operation (on page 3-23)
:TRACe:MAKE (on page 6-165)
:TRACe:DELete
This command deletes a user-defined reading buffer.
Type Affected by Where saved Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:TRACe:DELete "<bufferName>"
<bufferName>
A string that contains the name of the user-defined reading buffer to delete
Details
You cannot delete the default reading buffers, defbuffer1 and defbuffer2.
Example
TRAC:DEL "testData"
Delete the testData buffer.
Also see
Reading buffers (on page 3-2)
Remote buffer operation (on page 3-23)
:TRACe:MAKE (on page 6-165)
:TRACe:FILL:MODE
This command determines if a reading buffer is filled continuously or is filled once and stops.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
defbuffer1: CONT
defbuffer2: CONT
User-defined buffers: ONCE
Usage
:TRACe:FILL:MODE <fillType>
:TRACe:FILL:MODE <fillType>, "<bufferName>"
:TRACe:FILL:MODE?
:TRACe:FILL:MODE? "<bufferName>"
<fillType>
Fill the buffer continuously: CONTinuous
Fill the buffer, then stop:
ONCE
<bufferName>
A string that indicates the reading buffer; the default buffers
(defbuffer1 or defbuffer2) or the name of a user-defined
buffer; if no buffer is specified, defbuffer1 is used

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals