Section
6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
6-4 2461-901-01 A/November 2015
:MEASure?
This command makes measurements, places them in a reading buffer, and returns the last reading.
Type Affected by Where saved Default value
Query only Not applicable Not applicable Not applicable
Usage
:MEASure?
:MEASure:<function>?
:MEASure:<function>? "<bufferName>"
:MEASure:<function>? "<bufferName>", <bufferElements>
:MEASure? "<bufferName>"
:MEASure? "<bufferName>", <bufferElements>
<function>
The measure function:
Current: CURRent[:DC]
Resistance: RESistance
Voltage:
VOLTage[:DC]
<bufferName>
The name of the buffer where the reading is stored; if nothing is specified,
defbuffer1
is used
<bufferElements>
See Details
Details
This command makes a measurement using the specified function and stores the reading in a
reading buffer.
If you do not define the function parameter, the instrument uses the presently selected measure
function.
This query makes the number of readings specified by [:SENSe[1]]:COUNt. When you use a
reading buffer with a command or action that makes multiple readings, all readings are available in
the reading buffer. However, only the last reading is returned as a reading with the command.
If you define a specific reading buffer, the reading buffer must exist before you make the
measurement.
To get multiple readings, use the :TRACe:DATA? command.
Sending this command changes the measurement function to the one specified by <function>.
This function remains selected after the measurement is complete.
:MEASure? performs the same function as READ?.
:MEASure:<function>? performs the same function as sending :SENse:FUNCtion, then READ?.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV