Section
8: TSP command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
8-282 2461-901-01 A/November 2015
Details
The branch-once-excluded block is ignored the first time the trigger model encounters it. After the first
encounter, the trigger model goes to the specified branching block.
The branch-once-excluded block is reset when the trigger model starts or is placed in idle.
Example
trigger.model.setblock(2, trigger.BLOCK_BRANCH_ONCE_EXCLUDED, 4)
When the trigger model reaches block 2 the first time, the trigger model goes to block 3. If the trigger model
reaches this block again, the trigger model goes to block 4.
Also see
Once excluded block (on page 3-121)
trigger.model.setblock() — trigger.BLOCK_BUFFER_CLEAR
This function defines a trigger model block that clears the reading buffer.
Type TSP-Link accessible Affected by Where saved Default value
Function
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
Not applicable
Usage
trigger.model.setblock(blockNumber, trigger.BLOCK_BUFFER_CLEAR)
trigger.model.setblock(blockNumber, trigger.BLOCK_BUFFER_CLEAR, bufferName)
blockNumber
The sequence of the block in the trigger model
bufferName
The name of the buffer, which must be an existing buffer; if no
buffer is defined,
defbuffer1
is used
Details
When trigger model execution reaches the buffer clear trigger block, the instrument empties the
specified reading buffer. The specified buffer can be the default buffer or a buffer that you defined.
If you are clearing a user-defined reading buffer, you must create the buffer before you define this
block.
Example
trigger.model.setblock(3, trigger.BLOCK_BUFFER_CLEAR, capTest2)
Assign trigger block 3 to buffer clear; when the trigger model reaches block 3, it clears the reading buffer
named
capTest2
.
Also see
buffer.make() (on page 8-14)
Buffer clear block (on page 3-108)

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV