Section
6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
6-190 2461-901-01 A/November 2015
:TRIGger:BLOCk:BRANch:EVENt
This command branches to a specified block when a specified trigger event occurs.
Type Affected by Where saved Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:TRIGger:BLOCk:BRANch:EVENt <blockNumber>, <event>, <branchToBlock>
<blockNumber>
The sequence of the block in the trigger model
<event>
The event that must occur before the trigger model branches the specified block
<branchToBlock>
The block number of the trigger model block to execute when the specified event
occurs
Details
The branch-on-event block goes to a branching block after a specified trigger event occurs. If the
trigger event has not yet occurred when the trigger model reaches the branch-on-event block, the
trigger model continues to execute the blocks in the normal sequence. After the trigger event occurs,
the next time the trigger model reaches the branch-on-event block, it goes to the branching block.
If you set the branch event to none, an error is generated when you run the trigger model.
The f oll owing table shows the constants for the events.
Trigger events
Event description Event constant
No trigger event
NONE
Front-panel TRIGGER key press
DISPlay
Notify trigger block <n> (1 to 8); the trigger model generates a
trigger event when it executes the notify block
NOTify<n>
A command interface trigger:
Any remote interface: *TRG
GPIB only: GET bus command
VXI-11: VXI-11 command device_trigger
COMMand
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line <n> (1 to 6)
DIGio<n>
Line edge detected on TSP-Link synchronization line <n> (1 to 3)
TSPLink<n>
Appropriate LXI trigger packet is received on LAN trigger object
<n>
(1 to 8)
LAN<n>
Trigger event blender <n> (1 or 2), which combines trigger events
BLENder<n>
Trigger timer <n> (1 to 4) expired
TIMer<n>
Source limit condition occurs
SLIMit

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV