6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
6-16 2461-901-01 A/November 2015
Example
Saves the instrument settings in memory location
2.
Also see
Saving setups (on page 2-137)
*RCL (on page 6-15)
ACAL subsystem
Automatic calibration removes measurement errors that are caused by the performance drift on the
components used in the source-measure unit (SMU) digitizer as a result of temperature and time.
If you are using the digitize functions on the Model 2461, it is important to run the automatic
calibration at least once a week.
:ACAL:COUNt?
This command returns the number of times automatic calibration has been run.
Type Affected by Where saved Default value
Usage
:ACAL:COUNt?
Details
The number of times that autocalibration has been run since the last factory calibration. The count
restarts at 1 after a factory calibration.
Example
Returns the number of times auto calibration has
been run.
Example output:
Also see
Autocalibration (on page 3-51)
:ACAL:RUN (on page 6-18)