2461-901-01 A/November 2015 8-91
Details
This function retrieves the value of a global variable from the run-time environment of this node.
Do not use this command to retrieve the value of a global variable from the local node. Instead,
access the global variable directly. This command should only be used from a remote master when
controlling this instrument over a TSP-Link
®
network.
Example
print(node[5].getglobal("test_val"))
Retrieves and outputs the value of the global variable
named
test_val
from node 5.
Also see
node[N].setglobal() (on page 8-91)
node[N].setglobal()
This function sets the value of a global variable.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
node[N].setglobal(name, value)
N
The node number of this instrument (1 to 64)
name
The global variable name to set
value
The value to assign to the variable
Details
From a remote node, use this function to assign the given value to a global variable.
Do not use this command to create or set the value of a global variable from the local node (set the
global variable directly instead). This command should only be used from a remote master when
controlling this instrument over a TSP-Link
®
.
Example
node[3].setglobal("x", 5)
Sets the global variable x on node 3 to the value of 5.
Also see
node[N].getglobal() (on page 8-90)

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV