Model 2461
Interactive SourceMeter® Instrument Reference Manual Section 3:
Functions and features
2461-901-01 A/November 2015 3-77
Instrument access
You can specify that the control interfaces request access before taking control of the instrument.
There are several modes of access.
You can set one of the following levels of access to the instrument:
Full: Allows full access for all users from all interfaces
Exclusive: Allows access by one remote interface at a time with logins required from other
interfaces
Protected: Allows access by one remote interface at a time with passwords required on all
interfaces
Lockout: Allows access by one interface (including the front panel) at a time with passwords
required on all interfaces
The front panel is read-only when you are using a remote interface. You can view information and
swipe screens without being prompted to leave remote mode. If you attempt to make a change from
the front panel while the instrument is controlled from a remote interface, you will be prompted to
enter a password to gain access.
When you set access to full, the instrument accepts commands from any interface with no passwords
required. You can change interfaces as needed.
When you set access to exclusive, you must log out of one remote interface and log into another one
to change interfaces. To use another interface, log out of the present interface before logging into the
new interface. You do not need a password with this access.
Protected access is similar to exclusive access, except that you must enter a password when logging
in.
When you set access to locked out, a password is required to change interfaces, including the front-
panel interface.
Changing the instrument access mode
To change the access mode from the front panel:
1. Press the MENU key.
2. Under System, select Settings. The SYSTEM SETTINGS menu opens.
3. Press the button next to Access Mode.
4. Select the level of password access control you want to enable.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals