EasyManuals Logo

Keithley 2461 Reference Manual

Keithley 2461
1109 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1022 background imageLoading...
Page #1022 background image
Model 2461
Interactive SourceMeter® Instrument Reference Manual Section 9:
Frequently asked questions (FAQs)
2461-901-01 A/November 2015 9-9
How do I save the present state of the instrument?
You can save the settings in the instrument as a script using the front-panel menus or from a remote
interface. After they are saved, you can recall the script or copy it to a USB flash drive.
From the front panel:
1. Configure the Model 2461 to the settings that you want to save.
2. Press the MENU key.
3. Under Scripts, select Create Setup. The CREATE SETUP window is displayed.
4. Select Create. A keyboard is displayed.
5. Use the keyboard to enter the name of the script.
6. Select the OK button on the displayed keyboard. The script is added to internal memory.
Using SCPI commands:
Configure the instrument to the settings that you want to save. To save the setup, send the
command:
*SAV <n>
Where <n> is an integer from 0 to 4.
In the front-panel script menus, the setups saved with the *SAV command have the name Setup0x,
where x is the value you set for <n>.
Using TSP commands:
Configure the instrument to the settings that you want to save. To save the setup, send the
command:
createconfigscript("setupName")
Where setupName is the name of the setup script that will be created.
Why did my settings change?
Many of the commands in the Model 2461 are saved with the source or measure function that was
active when you set them. For example, assume you have the measure function set to current and
you set a value for display digits. When you change the measure function to voltage, the display digits
value changes to the value that was last set for the voltage measure function. When you return to the
current measure function, the display digits value returns to the value you set previously.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV