Interactive SourceMeter® Instrument Reference Manual Appendix C:
Status model
2461-901-01 A/November 2015 C-17
Clearing registers
Registers in the status model can be cleared using commands or by instrument actions. When a
register is cleared, the bits in the register are set to 0.
The event log and all registers are cleared when instrument power is cycled.
When you read a bit from the Operation Event, Questionable Event, or Standard Event Status
Register, the entire 16-bit or 8-bit register value is returned. The event register is cleared or set to 0.
Using SCPI commands:
To clear the event registers of the Questionable Event Status Register and Operation Event Status
Register sets, Standard Event Register, and Status Byte Register, send:
*CLS
When using the SCPI interface, this command does not affect the Questionable Event Enable
Register and Operation Event Enable Register sets.
To clear the Questionable Event Status Register, the Operation Event Status Register sets, Standard
Event Register, and Status Byte Register, send:
STATus:CLEar
When using the SCPI interface, this command does not affect the Questionable Event Enable
Register or Operation Event Enable Register sets.
Using TSP commands:
To clear the event registers of the Questionable Event Status Register and Operation Event Status
Register sets, Standard Event Register, and Status Byte Register send:
*CLS
To clear the Questionable Event Status Register, the Operation Event Status Register sets, Standard
Event Register, and Status Byte Register send:
status.clear()
When using the SCPI interface, this command does not affect the Questionable Event Enable
Register or Operation Event Enable Register sets.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV