Section
8: TSP command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
8-236 2461-901-01 A/November 2015
trigger.blender[N].clear()
This function clears the blender event detector and resets the overrun indicator of blender N.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.blender[N].clear()
N
The blender number (1 or 2)
Details
This command sets the blender event detector to the undetected state and resets the overrun
indicator of the event detector.
Example
trigger.blender[2].clear()
Clears the event detector for blender 2.
Also see
None
trigger.blender[N].orenable
This attribute selects whether the blender performs OR operations or AND operations.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Trigger blender N reset
Configuration script
false (AND)
Usage
orenable = trigger.blender[N].orenable
trigger.blender[N].orenable = orenable
orenable
The type of operation:
true: OR operation
false
: AND operation
N
The blender number (1 or 2)
Details
This command selects whether the blender waits for any one event (OR) or waits for all selected
events (AND) before signaling an output event.