Model 2461
Interactive SourceMeter® Instrument Reference Manual Section 3:
Functions and features
2461-901-01 A/November 2015 3-107
The programming example below illustrates how to clear triggers, turn on the SMU output, and then
enable a 30-second timeout to wait for a command interface trigger. When the trigger is received, the
instrument performs a voltage reading.
-- Clear any previously detected command interface triggers.
trigger.clear()
-- Turn on output.
smu.source.output = smu.ON
-- Wait 30 seconds for a command interface trigger.
triggered = trigger.wait(30)
-- Get voltage reading.
reading = smu.measure.read()
-- Send command interface trigger to trigger the measurement.
*TRG
*TRG cannot be used in a script. It must be sent from a remotely connected host computer.
Trigger model
The trigger model controls the sequence in which source and measure actions occur. The Model
2461 trigger model is flexible, allowing you to control as much or as little as needed for your
measurement application.
When you are setting up a trigger model, you can choose the following options:
Wait for an event to occur before making another measurement
Notify other equipment and timers that an event has occurred
Wait for another piece of equipment to signal completion
Use measure configuration lists to apply different measure settings dynamically during trigger
model operation
Specify delays between events and measurements
Use source configuration lists to sweep source settings and values
Turn the source on and off with programmable delays to create pulses
Store measurements into a given buffer until an event occurs, then switch to another buffer
Conditionally take actions based on whether the measurement falls within set limits
Additional options are detailed in the following sections.
The Model 2461 includes predefined trigger models to allow you to quickly implement a trigger model.
You can also set up your own trigger models.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV