Section
6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
6-12 2461-901-01 A/November 2015
:READ:DIGitize?
This query makes a digitize measurement, places it in a reading buffer, and returns the latest reading.
Type Affected by Where saved Default value
Query only Not applicable Not applicable Not applicable
Usage
:READ:DIGitize?
:READ:DIGitize? "<bufferName>"
:READ:DIGitize? "<bufferName>", <bufferElements>
<bufferName>
The name of the buffer where the reading is stored; if nothing is specified,
defbuffer1 is used
<bufferElements>
See Details; if nothing is specified, READing is used
Details
You must set the instrument to a digitize function before sending this command.
This query makes the number of readings specified by [:SENSe[1]]:DIGitize:COUNt. If multiple
readings are made, all readings are available in the reading buffer. However, only the last reading is
returned as a reading with the command. To get multiple readings, use the :TRACe:DATA?
command.
When specifying buffer elements, you can:
Specify buffer elements in any order.
Include any or all of the buffer elements listed below in a single list. You can repeat elements as
long as the number of elements in the list is less than 14.
Use a comma to delineate multiple elements for a data point.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals