FORMat subsystem
The commands for this subsystem select the data format that is used to transfer instrument readings
over the remote interface.
:FORMat:ASCii:PRECision
This command sets the precision (number of digits) for all numbers returned in the ASCII format.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0
Usage
:FORMat:ASCii:PRECision <n>
:FORMat:ASCii:PRECision DEFault
:FORMat:ASCii:PRECision MINimum
:FORMat:ASCii:PRECision MAXimum
:FORMat:ASCii:PRECision?
:FORMat:ASCii:PRECision? DEFault
:FORMat:ASCii:PRECision? MINimum
:FORMat:ASCii:PRECision? MAXimum
<n>
The precision:
Automatic: 0
Specific value: 1 to 16
Details
This attribute specifies the precision (number of digits) for queries.
Note that the precision is the number of significant digits. There is always one digit to the left of the
decimal point; be sure to include this digit when setting the precision.
Example
:FORM:ASC:PREC 10
Set a precision of 10 digits. An example of the
output is:
-6.999999881E-01
Also see
:FORMat[:DATA] (on page 6-46)

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV