EasyManuals Logo

Keithley 2461 Reference Manual

Keithley 2461
1109 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #692 background imageLoading...
Page #692 background image
Model 2461
Interactive SourceMeter® Instrument Reference Manual Section 8:
TSP command reference
2461-901-01 A/November 2015 8-19
bufferVar.clear()
This function clears all readings and statistics from the specified buffer.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
bufferVar.clear()
bufferVar
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2) or a user-defined buffer
Example
reset()
testData = buffer.make(50)
trigger.model.load("SimpleLoop", 3, 0, testData)
trigger.model.initiate()
waitcomplete()
printbuffer(1, testData.n, testData)
testData.clear()
print("Readings in buffer after clear ="
.. testData.n)
trigger.model.initiate()
waitcomplete()
printbuffer(1, testData.n, testData)
Create a reading buffer named
testData, make three readings and
store them in testData, and then view
the readings.
Print number of readings in testData.
Output:
-4.5010112303956e-10, -
3.9923108222095e-12, -
4.5013931471161e-10
Clear the readings in testData.
Verify that there are no readings in
testData.
Output:
Readings in buffer after
clear = 0
Store three new readings in testData
and view those when complete.
Output:
4.923509754e-07,
3.332266330e-07,
3.974883867e-07
Also see
buffer.delete() (on page 8-11)
buffer.make() (on page 8-14)
bufferVar.clear() (on page 8-19)
print() (on page 8-92)
printbuffer() (on page 8-93)
Reading buffers (on page 3-2)
Remote buffer operation (on page 3-23)

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals