6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
6-228 2461-901-01 A/November 2015
:TRIGger:LOAD "LogicTrigger"
This command loads a predefined trigger model configuration that sets up a digital trigger through the digital I/O.
Type Affected by Where saved Default value
Command only Not applicable Not applicable Not applicable
Usage
:TRIGger:LOAD "LogicTrigger", <digInLine>, <digOutLine>, <count>
:TRIGger:LOAD "LogicTrigger", <digInLine>, <digOutLine>, <count>, <clear>
:TRIGger:LOAD "LogicTrigger", <digInLine>, <digOutLine>, <count>, <clear>, <delay>
:TRIGger:LOAD "LogicTrigger", <digInLine>, <digOutLine>, <count>, <clear>, <delay>,
"<bufferName>"
:TRIGger:LOAD "LogicTrigger", <digInLine>, <digOutLine>, <count>, <clear>, <delay>,
"<bufferName>", <readingBlock>
The digital input line (1 to 6); also the event that the trigger model will wait
The digital output line (1 to 6)
The number of measurements the instrument will make
To clear previously detected trigger events when entering the wait block:
ENTer
To immediately act on any previously detected triggers and not clear them
The delay time before each measurement (167 ns to 10 ks); default is 0 for
no delay
The name of the reading buffer, which may be a default buffer
(defbuffer1 or defbuffer2) or a user-defined buffer; default is
defbuffer1
Define a measure or digitize block for the trigger model; options are:
• ACTive: Add a measure or digitize block to the trigger model based
on the active function; if no option defined, ACTive is used
• MEASure: Adds a measure block to the trigger model
•
: Adds a digitize block to the trigger model
Details
This trigger model waits for a digital input event to occur, makes a measurement, and issues a notify
event. The notify event asserts a digital output line.
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel MENU key and under Trigger, selecting Configure. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
TRIGger:BLOCk:LIST? command to view the trigger model blocks in a list format.
This command replaces the :TRIGger:LOAD:TRIGger:EXTernal command, which is deprecated.