Interactive SourceMeter® Instrument Reference Manual Section 2:
General operation
2461-901-01 A/November 2015 2-137
Saving setups
You can save the present settings and any configuration lists that you have defined for the Model
2461 to internal memory or an external USB flash drive.
After the settings are saved, you can recall the settings. You can also set them to be the default
settings when the instrument is powered on.
If you are using TSP commands, saved setups are scripts and can be added, modified, and deleted
like any other script. See Introduction to TSP operation (on page 7-1, on page 7-1) for additional
information about working with scripts.
Save a user setup to internal memory
From the front panel:
1. Configure the Model 2461 to the settings that you want to save.
2. Press the MENU key.
3. Under Scripts, select Create Setup. The CREATE SETUP window is displayed.
4. Select Create. A keyboard is displayed.
5. Use the keyboard to enter the name of the script.
6. Select the OK button on the displayed keyboard. The script is added to internal memory.
Using SCPI commands:
Configure the instrument to the settings that you want to save. To save the setup, send the
command:
*SAV <n>
Where <n> is an integer from 0 to 4.
In the front-panel script menus, the setups saved with the *SAV command have the name Setup0x,
where x is the value you set for <n>.
Using TSP commands:
Configure the instrument to the settings that you want to save. To save the setup, send the
command:
createconfigscript("setupName")
Where setupName is the name of the setup script that will be created.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV