Interactive SourceMeter® Instrument Reference Manual Section 3:
2461-901-01 A/November 2015 3-73
Pulse measurements
If the Model 2461 is set to any measure function except digitize, the measurement occurs after the
pulse level has settled and before the transition to the bias level.
Figure 127: Pulse measurement timing
When setting up pulse measurements, configure your measurement settings first, before the pulse
settings. The Model 2461 uses the most recently specified measure settings for the pulse-measure
sweep.