Section
8: TSP command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
8-190 2461-901-01 A/November 2015
smu.source.pulse.level
This attribute selects a fixed amplitude for the selected source function when pulsing.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Recall settings
Instrument reset
Power cycle
Source configuration list
Save settings
Source configuration list
0
Usage
value = smu.source.pulse.level
smu.source.pulse.level = value
value
The source pulse level to set:
Current: 10.5 A to 10.5 A
Voltage:
105 V to 105 V
Details
This command sets the output level of the voltage or current source pulse when pulsing.
The sign of the source level dictates the polarity of the source. Positive values generate positive
voltage or current from the high terminal of the source relative to the low terminal. Negative values
generate negative voltage or current from the high terminal of the source relative to the low terminal.
If a manual source range is selected, the level cannot exceed the specified range. For example, if the
voltage source is on the 2 V range, you cannot set the voltage source level to 3 V. When auto range
is selected, the amplitude can be set to any level supported by the instrument because it will select
the correct range automatically when pulsing.
Use this command with a source configuration list to build a custom pulse sweep list that has unique
pulse level points. Then use that custom source configuration list as a parameter to the
smu.source.pulsesweeplist() command to configure the instrument for the pulse sweep.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals