Interactive SourceMeter® Instrument Reference Manual Section 6:
2461-901-01 A/November 2015 6-153
Details
The address can be set to any address value from 1 to 30. However, the address must be unique in
the system. It cannot conflict with an address that is assigned to another instrument or to the GPIB
controller.
A new GPIB address takes effect when the command to change it is processed. If there are response
messages in the output queue when this command is processed, they must be read at the new
address.
If command messages are being queued (sent before this command has executed), the new settings
may take effect in the middle of a subsequent command message, so care should be exercised when
setting this attribute from the GPIB interface.
You should allow ample time for the command to be processed before attempting to communicate
with the instrument again. After sending this command, make sure to use the new address to
communicate with the instrument.
*RST does not affect the GPIB address.
Example
:SYSTem:GPIB:ADDRess?
Sets the GPIB address and reads the
address.
Output:
Also see
GPIB setup (on page 2-57)
:SYSTem:LFRequency?
This query contains the power line frequency setting that is used for NPLC calculations.
Type Affected by Where saved Default value
Usage
:SYSTem:LFRequency?
Details
The instrument automatically detects the power line frequency (either 50 Hz or 60 Hz) when the
instrument is powered on.
Example
Check the line frequency.
Also see
None