Model 2461
Interactive SourceMeter® Instrument Reference Manual Section 8:
TSP command reference
2461-901-01 A/November 2015 8-31
bufferVar.sourcestatuses
This attribute contains the source status conditions of the instrument for the reading point.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes
Restore configuration
Instrument reset
Power cycle
Not applicable Not applicable
Usage
statusInformation = bufferVar.sourcestatuses[N]
statusInformation
The status value when reading N of the specified buffer was acquired;
Source status values are:
Overvoltage protection was active: buffer.STAT_PROTECTION
Measured source value was read: buffer.STAT_READBACK
Overtemperature condition was active: buffer.STAT_OVER_TEMP
Source function level was limited: buffer.STAT_LIMIT
Four-wire sense was used: buffer.STAT_SENSE
Output was on:
buffer.STAT_OUTPUT
bufferVar
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2
) or a user-defined buffer
N
The reading number N; can be any value from 1 to the number of readings in the
buffer; use the bufferVar.n command to determine the number of readings in
the buffer
Details
This buffer recall attribute holds an array (a Lua table) of the status values for all the readings in the
buffer. The status values are floating-point numbers that encode the status value for each
measurement taken; see the following tables for values.
Buffer status bits for source function when reading was acquired
Bit Name Decimal value Description
2
STAT_PROTECTION
4
Overvoltage protection was active
3
STAT_READBACK
8 Measured source value was read
4
STAT_OVER_TEMP
16
Overtemperature condition was active
5
STAT_LIMIT
32
Source function level was limited
6
STAT_SENSE
64
Four-wire sense was used
7
STAT_OUTPUT
128
Output was on

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV