Model 2461
Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2461-901-01 A/November 2015 6-145
:SYSTem:CLEar
This command clears the event log.
Type Affected by Where saved Default value
Command only Not applicable Not applicable Not applicable
Usage
:SYSTem:CLEar
Details
This command removes all events from the event log, including entries in the front-panel event log.
Also see
:SYSTem:ERRor[:NEXT]? (on page 6-147)
:SYSTem:COMMunication:LAN:CONFigure
This command specifies the LAN configuration for the instrument.
Type Affected by Where saved Default value
Command and query
Rear panel LAN reset
Nonvolatile memory
AUTO
Usage
:SYSTem:COMMunication:LAN:CONFigure "AUTO"
:SYSTem:COMMunication:LAN:CONFigure "MANual,<IPaddress>"
:SYSTem:COMMunication:LAN:CONFigure "MANual,<IPaddress>,<NETmask>"
:SYSTem:COMMunication:LAN:CONFigure "MANual,<IPaddress>,<NETmask>,<GATeway>"
:SYSTem:COMMunication:LAN:CONFigure?
AUTO
Use automatically configured LAN settings (default)
MANual
Use manually configured LAN settings
<IPaddress>
LAN IP address; must be a string specifying the IP address in dotted decimal
notation; required if the mode is set to manual (default "0.0.0.0")
<NETmask>
The LAN subnet mask; must be a string in dotted decimal notation (default
"255.255.255.0")
<GATeway>
The LAN default gateway; must be a string in dotted decimal notation (default
"0.0.0.0")