Model 2461
Interactive SourceMeter® Instrument Reference Manual Section 8:
TSP command reference
2461-901-01 A/November 2015 8-247
trigger.digout[N].stimulus
This attribute selects the event that causes a trigger to be asserted on the digital output line.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Digital I/O trigger N reset
Configuration script trigger.EVENT_NONE
Usage
event = trigger.digout[N].stimulus
trigger.digout[N].stimulus = event
event
The event to use as a stimulus; see Details
N
Digital I/O trigger line (1 to 6)
Details
The digital trigger pulsewidth command determines how long the trigger is asserted.
The trigger stimulus for a digital I/O line can be set to one of the trigger events that are described in
the following table.
Trigger events
Event description Event constant
No trigger event
trigger.EVENT_NONE
Front-panel TRIGGER key press
trigger.EVENT_DISPLAY
Notify trigger block N (1 to 8) generates a trigger event when
the trigger model executes it
trigger.EVENT_NOTIFYN
A command interface trigger:
Any remote interface: *TRG
GPIB only: GET bus command
VXI-11: VXI-11 command
device_trigger
trigger.EVENT_COMMAND
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line N
(1 to 6)
trigger.EVENT_DIGION
Line edge detected on TSP-Link synchronization line N
(1 to 3)
trigger.EVENT_TSPLINKN
Appropriate LXI trigger packet is received on LAN trigger
object N (1 to 8)
trigger.EVENT_LANN

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals