2: General operation Model 2461 Interactive SourceMeter® Instrument
Reference Manual
2-136 2461-901-01 A/November 2015
Setting up the Histogram
To set up the Histogram:
1. Press the MENU key.
2. In the View menu, select Histogram.
3. Select the Data tab.
4. Select the Bin Buffer. This reading buffer contains the data that is binned.
5. Select the Scale tab.
6. Set the Minimum Boundary to the lowest value of data to be binned. Any data below this value
is binned in the low outlier bin.
7. Set the Maximum Boundary to the highest value of data to be binned. Any data above this value
is binned in the high outlier bin.
8. Set the Number of Bins to the bins in which to group the data. Two additional outlier bins are
added to the number of bins to capture data that is outside the specified boundaries.
9. In most cases, set Method to SmartScale. SmartScale selects either the Auto Bin or Fit method
whichever is most appropriate. Fit adjusts the y-axis so that the tops of all bins are visible; Auto
Bin redistributes the data evenly in the bins based on present minimum and maximum
boundaries.
You can pinch and zoom to change the scale of data. When you adjust the data on the screen, Auto
Scale is turned off. You can turn it on in the Scale tab.
10. To clear the Histogram, on the Data tab, select Clear Buffer. This clears the data from the
selected reading buffer and the statistics for the buffer, which in turn clears the Histogram.
An example of a histogram tab set for 25 bins is shown in the following figure.
Figure 82: Model 2461 Histogram

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals