6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
6-84 2461-901-01 A/November 2015
Details
This command creates an empty configuration list. To add configuration indexes to this list, you need
to use the store command.
Configuration lists are not saved when the instrument is turned off. If you want to save a configuration
list, use a saved setup to store the instrument settings, which include defined configuration lists.
Example
:SOUR:CONF:LIST:CRE "MySourceList"
Creates a source configuration list named MySourceList.
Also see
Configuration lists (on page 3-30)
*SAV (on page 6-15)
:SOURce[1]:CONFiguration:LIST:STORe (on page 6-88)
:SOURce[1]:CONFiguration:LIST:DELete
This command deletes a source configuration list.
Type Affected by Where saved Default value
Command only Not applicable Not applicable Not applicable
Usage
:SOURce[1]:CONFiguration:LIST:DELete "<name>"
:SOURce[1]:CONFiguration:LIST:DELete "<name>", <index>
<name>
A string that represents the name of a source configuration list
<index>
A number that defines a specific configuration index in the configuration list
Details
Deletes a configuration list. If the index is not specified, the entire configuration list is deleted. If the
index is specified, only the specified configuration index in the list is deleted.
When an index is deleted from a configuration list, the index numbers of the following indexes are
shifted up by one. For example, if you have a configuration list with 10 indexes and you delete index
3, the index that was numbered 4 becomes index 3, and the all the following indexes are renumbered
in sequence to index 9. Because of this, if you want to delete several nonconsecutive indexes in a
configuration list, it is best to delete the higher numbered index first, then the next lower index, and so
on. This also means that if you want to delete all the indexes in a configuration list, you must delete
index 1 repeatedly until all indexes have been removed.
Example
:SOURce:CONF:LIST:DEL "MySourceList"
Deletes a configuration list named
MySourceList
.
:SOURce:CONF:LIST:DEL "MySourceList", 2
Deletes configuration index 2 in the
configuration list named
MySourceList
.
Also see
Configuration lists (on page 3-30)
:SOURce[1]:CONFiguration:LIST:CREate (on page 6-83)

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals