EasyManuals Logo

Keithley 2461 Reference Manual

Keithley 2461
1109 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #741 background imageLoading...
Page #741 background image
Section
8: TSP command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
8-68 2461-901-01 A/November 2015
eventlog.clear()
This function clears the event log.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
eventlog.clear()
Details
This command removes all events from the event log, including entries in the front-panel event log.
Also see
eventlog.next() (on page 8-69)
eventlog.save() (on page 8-71)
Using the event log (on page 2-142)
eventlog.getcount()
This function returns the number of unread events in the event log.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
eventlog.getcount()
eventlog.getcount(eventType)
eventType
Limits the return to specific event log types; set a cumulative integer value that
represents the event log types to:
Errors only: eventlog.SEV_ERROR or 1
Warnings only: eventlog.SEV_WARN or 2
Errors and warnings only: eventlog.SEV_WARN|eventlog.SEV_ERROR or
3
Information only: eventlog.SEV_INFO or 4
Errors and information only: eventlog.SEV_INFO|eventlog.SEV_ERROR
or 5
Warnings and information only:
eventlog.SEV_INFO|eventlog.SEV_WARN or 6
All events:
eventlog.SEV_ALL
or
7
Details
A count finds the number of unread events in the event log. You can specify the event types to return,
or return the count for all events.
This command reports the number of events that have occurred since the command was last sent or
since the event log was last cleared.
Events are read automatically when localnode.showevents is enabled. You can also read them
individually with eventlog.next().

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals