8: TSP command reference Model 2461 Interactive SourceMeter® Instrument
8-194 2461-901-01 A/November 2015
smu.source.pulsesweeplinear()
This function sets up a linear pulse sweep for a fixed number of pulse points.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
smu.source.pulsesweeplinear(configListName, biasLevel, start, stop, points,
pulseWidth)
smu.source.pulsesweeplinear(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable)
smu.source.pulsesweeplinear(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName)
smu.source.pulsesweeplinear(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay)
smu.source.pulsesweeplinear(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime)
smu.source.pulsesweeplinear(configListName, biasLevel, start, stop, points,
pulseWidth, MeasEnable, bufferName, sDelay, offTime, count)
smu.source.pulsesweeplinear(configListName, biasLevel, start, stop, points,
pulseWidth, MeasEnable, bufferName, sDelay, offTime, count, xBiasLimit)
smu.source.pulsesweeplinear(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit)
smu.source.pulsesweeplinear(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit, failAbort)
smu.source.pulsesweeplinear(configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit, failAbort, dual)
The name of the source configuration list that the sweep uses; this must be defined
before sending this command
Output level the instrument sources before the first pulse and returns to between
pulses (delay and off time):
• Current: -7.35 A to 7.35 A
•
The voltage or current source level at which the pulse sweep starts:
• Current: -10.5 A to 10.5 A
•
The voltage or current source level at which the pulse sweep stops:
• Current: -10.5 A to 10.5 A
•
The number of pulse-measure points between the start and stop values of the pulse
sweep (2 to 1e6)
The time at the amplitude level for each pulse:
• Extended operating area: 10 A at 100 V, 5 % duty cycle, maximum pulse
width: 1 ms, minimum pulse width 150 µs (load dependent)
• Normal operating area (DC): 99.99 % duty cycle, maximum pulse width
10,000 s, minimum pulse width 150 µs
Enable or disable measurements at the top of each pulse:
• Enable: smu.ON (default)
• Disable: