6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
6-142 2461-901-01 A/November 2015
:SYSTem:CCHeck?
This command indicates whether one or more connections failed the contact check operation.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable Not applicable
Usage
:SYStem:CCHeck?
Details
Use this query to determine if any connections in the system failed the contact check operation.
This command returns 0 if one or more connections exceed the threshold resistance level set by the
:SYSTem:CCHeck:THReshold command, or it returns 1 if no connections exceed that value.
If you get a failed indication when sending this command, you can use the :SYSTem:CCHeck:ALL?
command to determine which connection failed the contact check test.
Sending this command when contact check is not enabled results in an error.
Example
:SYSTem:CCheck?
Query whether any connections failed the
contact check operation.
Output:
1
Indicates that all connections passed the contact
check operation.
Also see
:SYStem:CCHeck:ALL? (on page 6-142)
:SYSTem:CCHeck:STATe (on page 6-143)
:SYSTem:CCHeck:THReshold (on page 6-144)
:SYSTem:CCHeck:ALL?
This query runs the contact check operation and returns the result of the test for high, low, and guard connections.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
SYSTem:CCHeck:ALL?
Details
Use this command to check the high, low, and guard connections to the device under test (DUT) for
excessive contact resistance.
For each of the three connections, this command returns 0 (false) if the threshold resistance exceeds
the level specified by the :SYSTem:CCHeck:THReshold command, or 1 (true) if it does not exceed
that value.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals