Page #1041 background imageLoading...
Page #1041 background image
Model 2461
Interactive SourceMeter® Instrument Reference Manual Appendix B:
Common commands
2461-901-01 A/November 2015 B-5
*IDN?
This command retrieves the identification string of the instrument.
Type Affected by Where saved Default value
Query only None Not applicable Not applicable
Usage
*IDN?
Details
The identification string includes the manufacturer, model number, serial number, and firmware
revision of the instrument. The string is formatted as follows:
KEITHLEY INSTRUMENTS,MODEL nnnn,xxxxxxxx,yyyyyy
Where:
nnnn is the model number
xxxxxxxx is the serial number
yyyyyy is the firmware revision level
Example
*IDN?
Output:
KEITHLEY INSTRUMENTS,MODEL 2461,01234567,1.0.0i
Also see
System information (on page 2-81)
*LANG
This command determines which command set is used by the instrument.
Type Affected by Where saved Default value
Command and query
Not applicable
Nonvolatile memory
SCPI
Usage
*LANG <commandSet>
*LANG?
<commandSet>
The command set to be used:
TSP
SCPI
Details
The remote command sets that are available include:
SCPI: An instrument-specific language built on the SCPI standard.
TSP: A scripting programming language that contains instrument-specific control commands that
can be executed from a stand-alone instrument. You can use TSP to send individual commands
or use it to combine commands into scripts.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV