EasyManuals Logo

Keithley 2461 Reference Manual

Keithley 2461
1109 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1098 background imageLoading...
Page #1098 background image
Appendix
D: Convert Model 2400 to Model 2461 code Model 2461 Interactive SourceMeter® Instrument
Reference Manual
D-26 2461-901-01 A/November 2015
Model 2400 command
:SYSTem:CCHeck
:SYSTem:CCHeck?
Model 2461 command
:SYSTem:CCHeck:STATe (on page 6-143)
:SYSTem:CCHeck:STATe?
Notes
Also refer to Contact check (on page 2-115).
Model 2400 command
:SYSTem:CCHeck:RESistance
:SYSTem:CCHeck:RESistance?
Model 2461 command
:SYSTem:CCHeck:THReshold (on page 6-144)
:SYSTem:CCHeck:THReshold?
Notes
Refer to Contact check (on page 2-115).
Model 2400 command
:SYSTem:CLEar
Model 2461 command
:SYSTem:CLEar (on page 6-145)
Notes
Model 2400 command
:SYSTem:ERRor:ALL?
Model 2461 command
Not available
Notes
See Using the event log (on page 2-142) for changes to error reporting.
Model 2400 command
:SYSTem:ERRor:CODE:ALL?
Model 2461 command
Not available
Notes
See Using the event log (on page 2-142) for changes to error reporting.
Model 2400 command
:SYSTem:ERRor:CODE[:NEXT]?
Model 2461 command
:SYSTem:ERRor:CODE[:NEXT]? (on page 6-148)
Notes
Returns only errors from the event log. See Using the event log (on page 2-
142) for changes to error reporting.
Model 2400 command
:SYSTem:ERRor:COUNt?
Model 2461 command
:SYSTem:ERRor:COUNt? (on page 6-148)
Notes
Returns only errors from the event log. See Using the event log (on page 2-
142) for changes to error reporting.
Model 2400 command
:SYSTem:ERRor[:NEXT]?
Model 2461 command
:SYSTem:ERRor[:NEXT]? (on page 6-147)
Notes
Returns only errors from the event log. See Using the event log (on page 2-
142) for changes to error reporting.
Model 2400 command
:SYSTem:GUARd
:SYSTem:GUARd?
Model 2461 command
Not available
Notes
Cable guard is the only option available on Model 2461.
Model 2400 command
:SYSTem:KEY
:SYSTem:KEY?
Model 2461 command
Not available
Notes

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals