Interactive SourceMeter® Instrument Reference Manual Section 3:
Functions and features
2461-901-01 A/November 2015 3-157
Group leader overview
Each group has a dynamic group leader. The last node in a group that performs any operation
initiated by the master node is the group leader.
The group leader:
Performs operations initiated by the master node
Initiates remote operations on any node with the same group number
Cannot initiate remote operations on any node with a different group number
Can use the waitcomplete() command without a parameter to wait for all overlapped
operations running on nodes in the same group
Assigning groups
Group numbers can range from zero (0) to 64. The default group number is 0. You can change the
group number at any time. You can also add or remove a node to or from a group at any time.
Each time the power for a node is turned off, the group number for that node changes to 0.
The following example code dynamically assigns a node to a group:
-- Assign node 3 to group 1.
node[3].tsplink.group = 1
Running test scripts and programs on remote nodes
You can send the execute() command from the master node to initiate a test script and Lua code
on a remote node. The execute() command places the remote node in the overlapped operation
state. As a test script runs on the remote node, the master node continues to process other
commands simultaneously.
Use the following code to send the execute() command for a remote node. The N parameter
represents the node number that runs the test script (replace N with the node number).
To set the global variable "setpoint" on node N to 2.5:
node[N].execute("setpoint = 2.5")
The following code demonstrates how to run a test script that is defined on the local node. For this
example, scriptVar is defined on the local node, which is the node that initiates the code to run on
the remote node. The local node must be the master node.
To run scriptVar on node N:
node[N].execute(scriptVar.source)
The programming example below demonstrates how to run a test script that is defined on a remote
node. For this example, scriptVar is defined on the remote node.
To run a script defined on the remote node:
node[N].execute("scriptVar()")
It is recommended that you copy large scripts to a remote node to improve system performance.

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV