6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument
6-112 2461-901-01 A/November 2015
Example
SOUR:FUNC VOLT
SOUR:VOLT:READ:BACK ON
DIG:FUNC "CURR"
DIG:CURR:RANG 1
DIG:CURR:SRAT 50000
SOUR:PULS:SWE:VOLT:LIN 0, 0, 70, 8, 2e-3, ON, "defbuffer1", 10e-3, 10e-3, 2, .1,
1, ON, ON
Reset the instrument to its default settings.
Set the source function to voltage and the source range to 100 V.
Set the digitize function to current and the digitize range to 1 A.
Set the digitizer sample rate to 50,000 readings per second.
Set up a linear pulse sweep to sweep from 0 V to 70 V and then back to 0 V in eight voltage steps. Set the
pulse width to 2 ms, the pulse delay to 10 ms and the off time to 10 ms. Set the bias current limit and pulse
current limit to 1 A.
Also see
Pulse operation (on page 3-64)
:SOURce[1]:SWEep:<function>:LIST (on page 6-128)
:SOURce[1]:PULSe:SWEep:<function>:LIST (on page 6-116)
Trigger model (on page 3-107)