Section
7: Introduction to TSP operation Model 2461 Interactive SourceMeter® Instrument
Reference Manual
7-10 2461-901-01 A/November 2015
Power up script
The power up script runs automatically when the instrument is powered on. To create a power up
script, save a new script and name it autoexec. The autoexec script is automatically saved to
nonvolatile memory. See Saving a script to nonvolatile memory (on page 7-8).
If an autoexec script already exists, you must delete it by sending the
script.delete("autoexec") command. Performing a system reset does not delete the
autoexec script.
To set up the power up script from the front panel:
1. Press the MENU key.
2. Under Scripts, select Run.
3. Select Copy to Power Up. A dialog box confirms that the script was copied.
4. Select OK.
To save the power up script using remote commands:
Send the command:
autoexec.save()
To delete the existing autoexec script, send the command:
script.delete("autoexec")
Commands that cannot be used in scripts
You cannot use the following commands as variables in scripts:
There are some functions that resemble some of the strings below, but are actually defined TSP
functions. For example, printbuffer() (on page 8-93) is a function you can use in scripts. If you are
uncertain, check the TSP command reference (on page 8-1) to verify that the string is part of a
defined function.
abort
login
bit
logout
createconfigscript
node
endflash
opc
endscript
prevflash
flash
printbuffer
fs
printnumber
io
scpi
loadscript
table
loadandrunscript
waitcomplete

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals