12E5119 Instruction Manual QuickTOC
®
ultra
IECEx 103
7 How to Work with the Analyser
7.7 Measurement Settings
LAR | PROCESS ANALYSERS AG
Number of flushes:
This value determines the number of flushes of the injection needle after a measurement. The injection
needle is rinsed after each measurement with 1 ml of rinsing water to allow good reproducibility. The
allowed values are between 1 and 5. The default value is 1.
Notice
The injection needle is rinsed once at the start of online operation.