12E5119 Instruction Manual QuickTOC
®
ultra
IECEx 233
9 Accessories and Options
9.5 CO2-Remover
LAR | PROCESS ANALYSERS AG
9.5.5 Care and Maintenance Actions (CO
2
-Remover)
Only minor effort is required to service and maintain the CO
2
-Remover. This section shows the best way
to look after your CO
2
-Remover to guarantee trouble-free operation. The documentation of maintenance
and service work is a precondition for any warranty and guarantee claims, and also represents a valuab-
le aid in locating resolutions when malfunctions occur (Chapter 10 from page 249).
Table 28: Visual Inspection Log (CO
2
-Remover)
Interval Visual Inspection Criteria OK Action
1 Day
Housing
no damage
no contamination
Contact Support
_____________________
Service-Lamp
red service-lamp is
inactiv
Contact Support
_____________________
Date: Signature:
Table 29: Care and Maintenance Log (CO
2
-Remover)
Interval Action
Kind of
Action
OK Comment
12,000
operating
hours
(or every 2
years)
Replace the desiccant cartridges Maintenance
Replace the internal ball valves Maintenance
Replace all seals which removed
during maintenance
Maintenance
24,000
operating
hours
(or every 4
years)
Replace the exhaust valves Maintenance
Date: Signature:
Notice
A corresponding copy of the log is in Chapter 13 from page 265.
If you have any question, please contact your local partner or the Technical
Support of LAR (Chapter 15 on page 283).