4 Installation
4.8 Rinsing Water
LAR | PROCESS ANALYSERS AG
4.8 Rinsing Water
Rinsing Water (provided by the operator) used for rinsing the injection needle is required for Start-Up.
See Chapter 6.1.2 on page 79.
4.9 Provide Acid Solution (TOC-Direct and TOC-Difference Method)
For the usage of the analyser with the TOC-Direct or TOC-Difference Method, an acid solution for strip-
ping out inorganic carbon compounds (TIC) must be provided. See Chapter 6.1.3 on page 80 for the
TOC-Difference Method or the Chapter 6.1.4 on page 80 for the TOC-Direct Method.
Warning
Warning of incorrect electric connections
The analyser may be damaged if it is not connected to the mains matching the
data stated on the nameplate.
Before the analyser is switched on, a check must be run on whether the local
voltage supply matches that on the nameplate.
If this power supply is not available, please contact your local partner or the
Technical Support of LAR (Chapter 15 on page 283).
Do not switch the analyser on before using it for the first time without the pre-
sence of a technician authorised by LAR - otherwise this voids the warranty of
your analyser.
Warning
Warning of incorrect setting-up
The analyser may be damaged if it is not connected to the mains matching the
data stated on the nameplate.
Do not switch the analyser on before using it for the first time without the pre-
sence of a technician authorised by LAR - otherwise this voids the warranty of
your analyser.
Notice
Disconnect the analyser from electric power prior to cabling the system.
Notice
For the start-up, please use an acid solution which is not older than five days.
It is best to keep canisters in the Reagent Cabinet (see Chapter 9.2 from
page 206).
!
!