162 Instruction Manual QuickTOC
®
ultra
IECEx 12E5119
8 Care and Maintenance
8.1 Overview of regular Care and Maintenance Actions LAR | PROCESS ANALYSERS AG
8.1 Overview of regular Care and Maintenance Actions
The following maintenance schedule provides an overview of recommended and regular actions for car-
ing and maintaining your analyser. Visual inspections are used to check the applicative need for care
and maintenance actions.
Table 17: Recommended, regular Care and Maintenance Actions
Interval Action
Kind of
Action
Page Reference
If neces-
sary
Replace silicon seals for glass
components
Maintenance
Chapter 8.7.17 on page 196
Replace seal for reactor foot Maintenance Chapter 8.7.18 on page 196
Replace gas filter Maintenance Chapter 8.7.21 on page 199
1 Week
Renew calibration standard Care Chapter 8.5.20 on page 179
Replace silicone seal for TIC-Port Maintenance Chapter 8.7.15 on page 195
1 Month
Clean needle guide seal Care Chapter 8.5.21 from page 179
Replace needle guide seal Maintenance Chapter 8.7.16 on page 196
3 Months
Check the furnace system Care Chapter 8.5.22 on page 180
Check the gas cooling pipes Care Chapter 8.5.23 on page 181
Clean and oil injection port Care Chapter 8.5.24 on page 182
Move pump tubes Care Chapter 8.5.25 on page 183
6 Months
Check the injection port Care Chapter 8.5.27 on page 184
Replace pump tubes Maintenance
Chapter 8.7.19 on page 198
and
Chapter 8.7.20 on page 199