198 Instruction Manual QuickTOC
®
ultra
IECEx 12E5119
8 Care and Maintenance
8.8 Function Tests LAR | PROCESS ANALYSERS AG
analyser.
Use the following log for the documentation of the function tests:
Table 21: Function test protocol
Visual Inspection Criteria OK Action
Analyser status • Zero signal is between
0 - 0.1 FSR
• Carrier Gas IN / OUT:
approx . l/h.
Contact support
_________________________
_________________________
_________________________
Tightness Test
• Carrier gas volume flow
falls to < 5 l/h
Contact Support
_________________________
Test Run (XY-System)
• All positions are hit by the
injection needle
Contact Support
_________________________
Checking the
Measurement Results
• Measurement results
matches the calibration
standard
Contact Support
_________________________
Checking the
Measurement Process
• Measurement Process
goes on without
complications
Contact Support
_________________________
Date: Signature: