228 Instruction Manual QuickTOC
®
ultra
IECEx 12E5119
9 Accessories and Options
9.4 Overflow Sampler LAR | PROCESS ANALYSERS AG
9.4.5 Perform Visual Inspection (Overflow Sampler)
Procedure of the Visual Inspection:
1. Obtain a protocol for visual inspection (overflow sampler) (copy from chapter 13.9 on page 289, if
necessary).
2. Follow the instructions below for visual inspections for each component.
3. Carry out care or maintenance measures if the test criteria have not been met to and document
them in the care and maintenance log (overflow sampler).
4. After inspection and any maintenance and service procedures (overflow sampler), complete the
documentation of the protocol (s).
9.4.5.1 Flow
Check at the see-through area of the lid of the overflow sampler if the sample flows in and out correctly.
9.4.6 Perform Care and Maintenance (Overflow Sampler)
Procedure for Care and Maintenance Measures:
1. Stop measuring operation of the analyser via the offline button.
2. Close the sample feed.
3. Disconnect all connections of the overflow sampler
4. Obtain a care and maintenance protocol (sample storage vessel) (copy from chapter 13.8 on page
289 if necessary).
5. Follow the instructions below for the care and maintenance for the component.
6. Document the actions taken in the care and maintenance log.
7. After the inspection and the care and maintenance, complete the documentation of the logs.
8. Restore all connections of the overflow sampler.
9. Open the sample feed.
10. Start measuring operation of the analyser via the online button.