12E5119 Instruction Manual QuickTOC
®
ultra
IECEx 217
9 Accessories and Options
9.2 Reagent Cabinet
LAR | PROCESS ANALYSERS AG
9.2.7.5 Replace Coarse Filter Unit
Recommendation: - Replace the coarse filter unit every 12 months.
1. To depressurise the filter unit, disconnect the compressor from the water separator (undo the tube
on the water separator).
2. Pull the black lug of the left container down and turn the container to remove it.
3. The white separator is visible.
4. Turn the black plastic disc and take the white separator.
5. Screw on the new filter.
6. Fit the container onto it.
7. Establish the connection to the compressor.
9.2.7.6 Replace Activated Carbon
Recommendation: - Replace the activated carbon every 12 months.