250 Instruction Manual QuickTOC
®
ultra
IECEx 12E5119
10 Minor disruptions - quickly solved
10.2 Breakdowns LAR | PROCESS ANALYSERS AG
10.2 Breakdowns
This section provides information on and solutions for possible malfunctions with the measurement
system. Possible causes and actions are listed in the following table - the problem type is used as an
indicator. If you have questions, please contact your local partner or the Technical Support of LAR
(Chapter 15 on page 283).
Table 32: Troubleshooting
Problem Possible cause Actions
Analyser does not start
after connecting the power
supply plug and switching
the main switch to „ON“
No voltage
No power
Check the fuses (Chapter 10.3 on
page 251).
Switch on if necessary.
Software does not start/
Screen is black
Main siwitch off
Blown fuse(s)
Switch on the analyser using the
main switch on the left side of the
analyser.
Check the fuses (Chapter 10.3 on
page 251).
Cooler temperature is too
low or high
Ambient temperature is
too low or high
Cooler fuses are blown
Cooler is out of order
Ensure the ambient temperature is
between 5 °C - 35 °C.
Replace the fuses
Contact your local partner or the
Technical Support of LAR
(Chapter 15 on page 283)
Furnace temperature is too
low or high
Furnace fuse is blown
Connecting cable is
defective
Furnace is defective
Switch on the fuses
Check the connections
Contact your local partner or the
Technical Support of LAR (v)
The difference of
temperature and
temperature regulator is
more than 10°C
Thermocouple is not at
the right position or
defective
Contact your local partner or the
Technical Support of LAR
(Chapter 15 on page 283)
Calibration values are not
plausible
Calibration standard
empty or prepared false
Produce a new calibration standard
Perform a new calibration
Measurement values sway
strongly
Sample vessels are
contaminated
High particle density
Bubbles in the injection
system
System leaky
Clean sample vessels
Check injection system (glass
syringe)
Check tighteness and rectify leaks
Measurement cannot start Temperature of cooler or
furnace is not in the
allowed range
Analyser is in
maintenance mode
Check the temperature of the cooler
and the furnace
Error E1810 is shown Emergency shutdown of
the furnace
Error in the position of
the injection system
Contact your local partner or the
Technical Support of LAR
(Chapter 15 on page 283)
immediately!