7 How to Work with the Analyser
7.7 Measurement Settings LAR | PROCESS ANALYSERS AG
7.7.1.1 Measurement Interval
Online Measurement:
This function is used to set the waiting time between measurements. The waiting time begins after com-
pletion of a measurement until the start of another measurement.
Check function:
This value determines at what intervals the analyser should perform automatic measurements and the-
reby self-check the functionality..
Remote Control:
This function can be used to turn the remote control on or off. With a digital 24 V signal, a measurement
is started.
Fig. 71: Measurement parameters .- Measurement interval (Example)
Notice
To activate the check function, contact the technical support of LAR
(Chapter 15.1 on page 283).
Notice
If the remote control is deactivated, no status (on / off) is displayed in the status
screen (Chapter 7.5 on page 97). If the remote control is activated, the "off"
status is displayed in the status screen (Chapter 7.5 on page 97) until a signal
is present (on).
The remote control is realized via the digital inputs of the analyser. See
Chapter 3.7.2.3 on page 40.