12E5119 Instruction Manual QuickTOC
®
ultra
IECEx 187
8 Care and Maintenance
8.7 Perform Maintenance Actions
LAR | PROCESS ANALYSERS AG
1. Undo the tube connectors of the teflon tube.
2. Remove the old teflon tube and replace it with a new one having the same length.
3. Attach the new teflon tube at the tube connectors provided.
8.7.4 Replace Reactor Pipe Filling / Reactor Pipe
Precondition: - Reactor Pipe Filling is blocked or Reactor Pipe is damaged.
Fig. 129: Reactor pipe filling (Standard)
Danger
Danger of burns
Allow the oven to cool completely (2 hours in total). One hour after switching
off the furnace, the temperature of the reactor tube is still about 400° C to
500° C!
Always use heat-resistant gloves when working on the stove.
Warning
Correct re-assembly of removed components
During assembly, remember that all parts that have been removed and / or di-
sassembled must be reassembled as they were in their original condition.
1 Reactor Pipe
2 Ceramic Protective Pipe
3 Ceramic Balls 1.6mm
4 Ceramic Balls 3.5 - 4.5mm
5 Ceramic Sieve
x
!