234 Instruction Manual QuickTOC
®
ultra
IECEx 12E5119
9 Accessories and Options
9.5 CO2-Remover LAR | PROCESS ANALYSERS AG
9.5.6 Perform Visual Inspections (CO
2
-Remover)
Procedure for visual inspections:
1. Use a visual inspection log (CO
2
-Remover) (Chapter 13.11 on page 276) for documentation.
2. Follow the instructions below for visual inspections of individual components.
3. Carry out care and/or maintenance actions when test criteria are not met, and document them in
the care and/or maintenance log (CO
2
-Remover).
4. After the check and any care and/or maintenance actions, conclude the documentation of the
log(s).
9.5.6.1 Housing
1. Check the housing of the CO
2
-Remover for integrity.
2. If you see damage, please contact Technical Support.
3. Check the CO
2
-Remover for cleanliness.
4. If contamination is detected, remove it with a damp cloth.
5. Once all test criteria are met, proceed with the next step.
9.5.6.2 Service Lamp
1. Check the red service light for inactivity.
2. If the red service light is lit, maintenance of the CO
2
-Remover is required to ensure the best
possible air quality. Please contact the Technical Support of LAR (Chapter 15 on page 283).
3. Once the test criterion is met, proceed with the next step.
Warning
Improper cleaning
Do not clean the CO
2
-Remover with abrasives or solvents. Use only a mild de-
tergents.
!