214 Instruction Manual QuickTOC
®
ultra
IECEx 12E5119
9 Accessories and Options
9.2 Reagent Cabinet LAR | PROCESS ANALYSERS AG
page 213).
9. Now fit the cover by hand and attach it using the four screws.
10. Fit the filter cartridge back into the pipe clamps.
11. Reconnect the upper and lower tubes to the filter cartridge.
9.2.7.2 Replace Gas Filter
Recommendation: - Replace the gas filter every 3 months.
1. Pull the connection tubes of the old filter from the connectors.
2. Connect the new filter with the connection tubes of the Ambient Air Preparation Unit, noting the
direction of flow (black arrow).
Fig. 138: Filter Cartridge (exploded assembly drawing)