12E5119 Instruction Manual QuickTOC
®
ultra
IECEx 223
9 Accessories and Options
9.3 Proceed analogously to the replacement of the soda lime in
LAR | PROCESS ANALYSERS AG
9.3.7 Perform Visual Inspections (FlowSampler
®
)
Procedure for visual inspections:
1. Use a visual inspection log (FlowSampler
®
) (Chapter 13.8 on page 273) for documentation.
2. Follow the instructions below for visual inspections of individual components.
3. Carry out care and/or maintenance actions when test criteria are not met, and document them in
the care and/or maintenance log (FlowSampler
®
).
4. After the check and any care and/or maintenance actions, conclude the documentation of the
log(s).
9.3.7.1 Flow
1. Look through the window to check if a steady, continuous liquid flow can be seen (5, Fig. 141,
page 219 or 5, Fig. 142, page 220).
2. Once the test criterion is met, proceed with the next step.
9.3.7.2 Sampling Pipe (Needle)
1. Check the cleanliness of the sampling pipe (needle) of the FlowSampler
®
.
2. Once the test criterion is met, proceed with the next step.