14 Instruction Manual QuickTOC
®
ultra
IECEx 12E5119
2 Operating Principle of the Analyser
2.7 Measurement and Working Ranges of the QuickTOC®ultra LAR | PROCESS ANALYSERS AG
2.7 Measurement and Working Ranges of the QuickTOC
®
ultra
The analyser can be deployed in several measurement ranges (application-specific). Each measure-
ment range is assigned a recommended working range. These working ranges are in the table below
Table 1: Overview of Measurement and Working Ranges
Measurement Ranges
[mg/l C]
Recommended Working Ranges
[mg/l C]
0 - 100 0.1 - 100
0 - 400 2 - 400
0 - 2,000 5 - 2,000
0 - 15,000 100 - 15,000
0 - 50,000 500 - 50,000